BRAND | Ntron |
Product | OC-25 -02-432 |
Description | NTR sensor |
Internal code | IMP3754104 |
Weight | 0.4 |
Technical specification | OC-25 (EC 0... 25%, NTRON thread) Electrochemical oxygen sensor for operation with NTRON oxygen analyzers Required for SIL2 in conjunction with the SILO2 transmitter Measuring range: 0... 25% O2 (overload: 30% max) Linearity: +- 0.2% of the measuring range Temperature range: - 20... +45 °C Pressure range: 900... 1115 mbar Response time (t95): <20 s, typically 12 s Electrical connection: 4-pin. connectors Case material: Delrin Certificates: ATEX II 1 DG Ex ia IIC T6 Ga, Ex ia IIIC T80oC Da - Gas and dust Ex Ntron oxygen sensor, model OC-25 Part Number: 02-432 ATEX/ IECEX Certified EEx ia IIC T6 ATEX II 1 GD Measurement range: 0-25% O2 Chemically resistant oxygen sensor Suitable for use on SIL 2 applications Process connection: Flow through or tri-clamp sensor base Electrical connection: IP67 connection. Product datasheet number: 02-432-P-001 |
default.send_request_amp_Group2